Our ophthalmic products

NIMO TR0815

The key to understanding refractive IOLs.

PICTURES:

PRODUCT DETAILS

Based on the patented “Phase Shifting Schlieren” (PSS) method, the NIMO TR0815 measures every type of refractive intraocular lenses, whether wet or dry, monofocal, toric, multifocal or multifocal toric.

Through its advanced technology, the NIMO offers the highest resolution amongst all commercially available power-mapping instruments, providing power and full wavefront characterization with matchless repeatability and reproducibility. Routine calibrations are no longer necessary and the software offers a user-friendly interface with plenty of options and full traceability.

Key characteristics:

  • Measures monofocal, toric, multifocal, multifocal toric lenses IOLs and more with the exact same hardware
  • Automatic marks detection for torics (TLMD) with assessment of the angle error between marks and cylinder axis
  • High-resolution power maps and power profiles
  • Rapid data acquisition – complete assessment of the most complex refractive IOLs in just a few seconds
  • Dynamic range -100D to +100D
  • Easy integration into automated production lines
  • One time calibration
  • Fully automatic Toric IOL measurement
  • Dry or wet measurement
  • Full customization program available

Hardware specifications:

Measurement principle Phase-shifting Schlieren in transmission mode
Field of view 8 x 8mm2
Clear aperture Maximum 8 mm diameter
Wavelength 546 nm
Imaging device 1024 x 1024 pixels
Spatial resolution 8 x 8 μm² or 16 x 16 μm² when Binning
Speed of measurement 4 to 10 seconds depending on ROI and measurement type
Absolute power range ± 70 D on 8 mm Ø and ± 100 D on 5 mm Ø
Repeatability Better than 0,005 D
Repeatability variance 0,01 D (ISO 5725)
External dimensions 170 x 220 x 590 mm³ (23 x 9 x 7 in³)

Software specifications:

Measurement conditions Wet or dry (polynomial correction available)
Spherical Sphere power, MTF (with or without ISO eye model), Zernike wavefront analysis
Toric Sphere power, cylinder, *axis, MTF in principal axis, Zernike wavefront analysis. *with optional TLMD plug-in
Aspheric Spherical aberration and MTF for 3 apertures, Zernike wavefront analysis
Multifocal Multi zone annular radial power profiles, base and add
Multifocal toric Multi zone sphere, cylinder, axis and radial power maps, base, add, cylinder
Modes of operation Production Batch or Single Measurements (full access to all options)
Data output Save image and graph as image (BMP), save profiles as text file (CSV). Batch report with one line per measurement, statistics, save projects